SP5 Zone 2
Type Fixed site sampler
Housing Double-walled stainless steel (material 1.4301/ SS304) with 40 mm insulation. Housing separated in sample compartment and control compartment, each with lockable door. Protective top which can be opened for connection and maintenance works. Option: material 1.4571/ SS316Ti or SS304/ SS316Ti -EPOXY coated
Thermo-Insulation Self-contained, controlled cooling / heating with 4 settings, no-frost. independent of the programmable controller, Temperature in sample compartment: 4°C (adjustable from 0,0-9,9°C)
Control Microprocessor control, Sleep-Mode (<5mA), power supply 8-16 V foil keyboard (with keys 0-9, ESC, ENT, cursor), graphical display (128*64 Pixel), back lit
Data logger 3000 entries, non-volatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals
Programming 12 freely programmable user programs, with function to link programs.
Program start options - IMMEDIATELY; - DATE/TIME - WEEKDAY/TIME; - BY AN EXTERNAL SIGNAL
Program End/Stop options End of sampling program - AFTER 1 RUN - AFTER X RUNS - CONTINUOUS OPERATION - DATE/TIME
Pause mode Interruption of program run at any time
Overfilling protection Adjustable from 1–999 samples/bottles
Interval setting 1 min. to 99 h 59 min. in steps of 1 minute
Sampling method -Vacuum system 20-350 ml Option: bypass system 20-250 ml (compressed air necessary) Explosion-proof diaphragm pump 230 V, vacuum 6 m, pressure 1 bar, change-over from pressure/vacuum/aeration by means of explosion-proof solenoid valves.
Pulse setting 1 to 9999 pulses/sample
Manual sample extraction Possible at any time without interrupting the current program run
Program protection Up to 5 years after voltage loss
Interface Not connected due to ATEX certification
Signal inputs • 2 x analogue: 0/4-20 mA, • 8 x digital (flow, event, 1 inputs can be programmed freely) option: expandable with 4x digital, 3 inputs can be programmed freely, and 8x analogue 0- 20 mA or 0-10 V, Impulslength 60ms, switching level 7-24 V, max. working restistance 500 Ohm, max. length of signalcable 30 m
Bottle variants Plastic 1 x 25 L, 1 x 50 L, 2 x 10 L 4 x 6,0 L, 4 x 10 L, 4 x 14 L 12 x 2,9 L Glass 12 x 2,0 L
Wetted materials PC, PVC, Silicone, PS, PE, EPDM (optional: metering vessel glass Duran50, sinker weight SS304)